Water Quality - Sensors
Robust, pluggable multi-parameter sensor made of plastic (PVC-U) for mounting up to twelve sensors (water level, temperature, conductivity, pH, redox potential, oxygen, turbidity, ammonia, nitrate, chloride, ammonium, sodium, calcium, fluoride, potassium, chlorophyll or cyano-bacteria-rhodamine WT)
Additional electric contact gauge function combined with KLL-Q-2 mobile field laboratory
Optimised design (Ø 89 mm, 572 mm length), usable starting from a 4"-diameter pipe
Output: RS485 (SHWP), Option: MODBUS, SDI-12, 0/4-20 mA
Optional: integrated, high-performance SEBA data logger Qualilog16 (for stationary use)
Individually configurable:
Depending on the measurement task at hand, the MPS-K16 can be individually equipped with a maximum of twelve water quality sensors and can detect up to 17 measurement parameters.
Simple handling:
Special bayonet locks grant the users a simple and quick replacement of worn sensors. The calibration and configuration of the MPS-K16 is done by means of a special USB-RS485 interface cable in connection with SEBAConfig operation software (Windows). This will automatically detect if a probe is spent and must be replaced.
Robust:
The housing of the MPS-K16 is made of durable plastic (PVC-U). The probe can thus be used even under adverse ambient conditions (e.g. landfill sites, contaminated sites).
Variable:
The proven and robust SEBA-connect feature of the MPS-K16 allows both the connection to mobile handsets (SEBA electric contact meter, KLL-Q-2, Checker-2) and stationary data collectors with (UniLogCom, LogCom/FlashCom) or without data transmission (Qualilog16).
Parameter: Water level [cm], [m]
Sensor:
Relative pressure probe
Measuring principle:
Capacitive with ceramic measurement cell
Measuring range: 0-2 / 1 /4/ 5/ 20/ 40/ 100 / 200m Water column or according to specification
Measuring accuracy: ± 0.1% FS
Long-term stability: ± 0.1% / year
Parameter: Water temperature [° C]
Sensor: NTC30 with polynomial linearization
Measuring Principle: NTC resistor
Measuring range: -5 - + 50 ° C
Measuring accuracy: 0.1 ° C
Resolution: 0.01 ° C
Pressure range: 0-50 bar
Parameter: Water temperature [° C]
Sensor: NTC30 with polynomial linearization
Measuring principle: NTC resistor
Measuring range: -5 - + 50 ° C
Measuring accuracy: 0.1 ° C
Resolution: 0.01 ° C
Pressure range: 0-50 bar
Parameter: Conductivity [μS / cm], [mS / cm]
Derived parameters: Solutes [g / l], salinity, density [g / l]
Sensor: 4-electrode measuring cell
Measuring range: 0-200mS / cm (total) automatic
Measuring range switching: 0-200μS / cm; 0.2-2mS / cm; 2-20 mS / cm; 20-200 mS / cm
Measuring accuracy: ± 1μS / cm at measuring range 0-200μS / cm; +/- 0.5% with measuring range 0.2-200mS
Resolution: 0.0001 mS / cm
Temperature: -5 ° C-50 ° C
Pressure range: 0-50 bar
Parameter: PH value [pH]
Sensor: Glass electrode
Measuring range: 0-14pH
Measuring accuracy: ± 0.1pH
Resolution: 0.01pH
Temperature: 0-50 ° C
Pressure range: 0-20 bar
Parameter: Redox [mV]
Sensor: Glass electrode
Measuring range: -1200mV-1200mV
Measuring accuracy: ± 10mV
Resolution: 0.1mV
Temperature: 0-50 ° C
Pressure range: 0-20 bar
Parameter: Oxygen [mg / l]
Derived parameters: Oxygen saturation [%]
Sensor: Clark electrode (amperometric)
Measuring range: 0-40mg
Measuring accuracy: ± 0.5% FS
Resolution: 0.01mg / l
Temperature: 0-50 ° C
Pressure range: 0-10 bar
Alternatively:
Sensor: Oxygen [mg / l]
Measuring range: 0-25mg / L (at 25 ° C, 1013hPa) 0-40mg / L (at 3 ° C, 1013hPa)
Measuring accuracy: ± 0.5% FS
Resolution: 0.001mg / l
Temperature: -5 ° C-50 ° C
Pressure range: 0-12 bar
Sensor: Turbidity [NTU]
Derived parameters: Suspended substances [mg / l] optically
Measuring range: 0-1000 NTU and 0-5000 NTU
Measuring accuracy: ± 1% of the respective end of measuring range
Resolution: ± 0.0125 % of the respective end of measuring range
Temperature: 0-50 ° C
Pressure range: 0-10bar (with wiper) 0-20bar (without wiper)
Further possible parameters for MPS-K16:
Housing:
Material: PVC-U
Dimension: Ø89
Length: 572mm
Protection class:
Probe: IP68
RS485 interface: IP67 (immersion depth with dry cartridge up to 3m, depending on cable length)
Pressure range: 0-12 bar
Outputs: RS485 (SHWP) Standard, MODBUS, SDI12, 4-20mA (option)
Communication Interface: RS485
Power supply: 4-15VDC or 8-15VDC with optical sensors
Operating temperature: Parameter-dependent
Mounting depth: Up to 1000m
Electric wire: Reinforced cable with pressure compensation line
Gore-tex membrane, dry cartridge with SilicaGel forthe protection of the pressure compensation line