System Introduction
Reasonable assessment of water balance and evapotranspiration is an important link in the hydrological cycle, and lysimetry system is a brand-new tool for assessing the utilization of water in soil; the storage and transfer of water in the same location. The micro lysimetry system is specially used to study the long-term monitoring of the hydrological cycle of the ecosystem. With the continuous in-depth development of soil physics, meteorology and other fields, it can be used for research on landfill, soil pollution, groundwater pollution and other related aspects.
The material of the soil column bucket is stainless steel, and the size can be customized according to the needs. For small soil column research, the recommended size is 30-60cm in diameter and 30-100cm in height (other sizes can be customized according to customer needs). soil or backfill for monitoring.
Measurement Parameters:
Soil moisture, soil temperature, soil evaporation, soil infiltration
Optional measurement parameters: soil conductivity, soil heat flux, soil water potential, soil solution sampling
System Features
It can display data in real time and receive data regularly.
Send data in real time and download data automatically and regularly, without the need for personnel to collect data.
The output data is in tabular format, which is easy to be called under Excel and various data processing software and database software
ST-100: Thermistor Temperature Sensor
SM150T Soil Moisture Sensor
Measurement | Volumetric water content and soil temperature |
Accuracy | ± 0.03 m3.m-3 (3%) With soil-specific calibration |
± 0.5°C (0 to 40°C for temp sensor) ± 0.75°C (-20 to +60°C for temp sensor) | |
Soil moisture measurement range | Full accuracy over: 0 to 0.7 m3.m-3 |
Full range : 0 to 1.0 m3.m-3 | |
Salinity range | 50 to 500 mS.m-1 |
Salinity errors < 0.035 m3.m-3 from 0.05 to 0.4 m3.m-3. Can be calibrated up to 2000 mS.m-1 | |
Temperature range | Full accuracy over: 0 to 40°C |
Output | 0 to 1.0 V differential Corresponding to 0 to ~0.6 m3.m-3 |
Resistance 5.8Ω to 28Ω for temp sensor | |
Power requirement | 5-14 V, ~18 mA for 1s |
Minimum 5 V with 100 m cable | |
Environmental | IP68 (-40 to +70°C) |
Sample volume | ~55 x 70 mm diameter |
Sample volume is weighted towards soil immediately surrounding the rods | |
Dimensions and weight | Overall: 143 x 40 mm diameter Rods: 51 mm x 2.5 mm diameter Weight: 0.1 kg (exc. cable) |
Sensor calibrations | Individual sensors are interchangeable Recalibration advised every 3 years (depending on use) |
Soil calibrations | Generalised Mineral and Organic soil calibrations are supplied |
MPS-6 soil water potential and temperature sensor
Soil Load Cell
Please confirm the system configuration and some key technical specifications with our company before purchasing.